Appendix A: Validated OpenArsenic method parameters

Summarizes the analytical parameters used during validation of the OpenArsenic platform.

Appendix A: Validated OpenArsenic method parameters

A.1 Purpose

This appendix summarizes the analytical parameters used during validation of the OpenArsenic platform.

Unless otherwise stated, users are encouraged to use these parameters for routine groundwater monitoring.

A.2 Hardware configuration

Component Specification
Potentiostat Rodeostat
Software OpenVoltam
Working Electrode Gold microwire
Gold Wire Diameter 30 µm
Exposed Gold Length ~5 mm
Reference Electrode Ag/AgCl (3 M KCl)
Counter Electrode Inert metal wire
Cell Volume 40 mL
Stirring Magnetic stirrer
Potentiostat Rodeostat
Software OpenVoltam
Working Electrode Gold microwire

A.3 Measurement parameters

Parameter Value
Supporting Electrolyte HCl
Final Electrolyte Concentration 0.1 M HCl
Deposition Potential -1.3 V
Deposition Time 20 s
Background Deposition Time 1 s
Start Potential -0.4 V
End Potential +0.7 V
Scan Rate 4 V/s
Sampling Frequency 480 Hz
Current Range 10 µA
Replicates per Run 4
Supporting Electrolyte HCl

A.4 Current validation status

Metric Current performance
Detection Limit ~0.5 ppb
Working Range ~0.5–25 ppb
Groundwater Samples Evaluated 11
ICP-MS Comparison Strong agreement
User Testing Ongoing
Day-to-day reproducibility 6.2% RSD
*based on 20 measurements of the same sample in 2 weeks