Chapter 12: Standard addition method

Instructions for the OpenArsenic standard addition scheme.

Chapter 12: Standard addition method

12.1 Introduction

The OpenArsenic platform uses the standard addition method to quantify dissolved arsenic concentrations in groundwater. Standard addition is a widely used analytical approach that is particularly useful when sample composition varies between locations and may influence analytical response.

Groundwater samples often contain different concentrations of dissolved salts, metals, organic matter, and other constituents. These differences can affect electrochemical measurements and make direct calibration more difficult. The standard addition method helps compensate for these matrix effects by performing calibration directly within the sample itself.

For this reason, standard addition forms a central part of the OpenArsenic analytical workflow.

12.2 Why standard addition?

Many analytical methods rely on external calibration standards. In an external calibration approach:

  1. A series of standards are prepared. 
  2. A calibration curve is generated. 
  3. Unknown samples are compared against that curve. 

While effective in many situations, external calibration assumes that standards and samples behave similarly during measurement. Groundwater samples can vary considerably in composition, potentially introducing matrix effects that influence analytical response. Standard addition minimizes these effects by adding known quantities of arsenic directly to the sample being analyzed.

As a result:

  • Calibration occurs within the sample matrix. 
  • Matrix effects are partially compensated. 
  • Quantification is generally more robust. 

12.3 The OpenArsenic standard addition scheme

The current OpenArsenic method uses:

Measurement 1: Sample

Measurement 2: Sample + 2.5 ppb arsenic

Measurement 3: Sample + 5 ppb arsenic

This workflow was selected during method development because it:

  • Produces clear signal increases 
  • Remains within the validated working range 
  • Provides reliable regressions for typical groundwater samples 
  • Minimizes reagent consumption 

12.4 Arsenic standard solution

The current OpenArsenic workflow uses: 1 ppm arsenic standard

This standard is used for all routine additions.

The standard should:

✔ Be prepared accurately

✔ Be clearly labeled

✔ Be protected from contamination

✔ Be replaced when quality is uncertain

The preparation of arsenic standards is described in Chapter 7.

12.5 What should happen during standard addition determination of arsenic

Following each standard addition, the arsenic peak should increase. Although replicate measurements will vary slightly, the overall trend should be clear.

In most successful measurements:

  • Peak position remains similar. 
  • Peak shape remains similar. 
  • Peak height increases. 

Figure 12.1. Example increase in arsenic signal following successive additions.

12.6 Arsenic calculation

After peak analysis is completed, OpenVoltam constructs a standard addition regression.

The regression uses:

  • Added arsenic concentration 
  • Measured analytical signal 

In the Sample Environment:

  1. Hold CTRL 
  2. Select: 
    • Sample 
    • Standard Addition 1 
    • Standard Addition 2 
  3. Click Calculate 

OpenVoltam currently supports several calculation approaches. In the screen that opens, select peak height over baseline.

Calculation approach Explanation Advantages Disadvantages
Peak height over baseline (Recommended) This method calculates peak height relative to the selected baseline.
  • Most robust
  • Best reproducibility
  • Best agreement with historical workflows
  • Recommended for routine arsenic analysis
User-biases with baseline definition
Peak height from zero Peak height is measured relative to the zero-current line. Simple More sensitive to baseline drift
Derivative methods:
  • Left-side derivative
  • Right-side derivative
  • Average derivative
Peak height is not used to calculate arsenic concentration but the maximum slope of left-side, right-side or average is taken.
  • No baseline needed
  • No user bias
  • Sensitive to noise and smoothing parameters
  • Sensitive to peak shape separation
  • Still considered experimental for the OpenArsenic system

The software then calculates the original arsenic concentration from the regression. For routine users, it is not necessary to perform these calculations manually. However, users should understand the basic principle.

Figure 12.2. Example regression generated from sample, +2.5 ppb, and +5 ppb measurements.

A successful regression should show:

✔ Increasing signal with increasing addition

✔ Approximately linear behavior

✔ No obvious outliers

If that is the case, select ‘Show results’ and the software will give the calculated arsenic concentration.

Figure 12.3: The calculated arsenic concentration

12.7 Warning signs

The following observations should prompt further investigation.

Observation Possible causes
No signal increase
  • Incorrect standard preparation
  • Incorrect addition volume
  • Analytical problems
Large scatter between replicates
  • Poor reproducibility
  • Electrode instability
  • Sample preparation errors
Non-linear regression
  • Incorrect additions
  • Peak analysis errors
  • Outside of working range
Large confidence intervals
  • Weak signal
  • Poor reproducibility
  • Outliers

12.8 Common mistakes

Mistake Recommendation
Incorrect standard concentration Always verify the concentration of the arsenic standard.
Incorrect addition volume Small pipetting errors can affect final calculations.
Poor documentation Always record sample characteristics and additions clearly to avoid issues later on.
Changing multiple variables Do not change:
  • Method parameters
  • Addition scheme
  • Electrode setup
within the same measurement series.
Ignoring failed regressions Poor regressions often indicate underlying analytical problems. Check your system.