Chapter 12: Standard addition method
Instructions for the OpenArsenic standard addition scheme.
12.1 Introduction
The OpenArsenic platform uses the standard addition method to quantify dissolved arsenic concentrations in groundwater. Standard addition is a widely used analytical approach that is particularly useful when sample composition varies between locations and may influence analytical response.
Groundwater samples often contain different concentrations of dissolved salts, metals, organic matter, and other constituents. These differences can affect electrochemical measurements and make direct calibration more difficult. The standard addition method helps compensate for these matrix effects by performing calibration directly within the sample itself.
For this reason, standard addition forms a central part of the OpenArsenic analytical workflow.
12.2 Why standard addition?
Many analytical methods rely on external calibration standards. In an external calibration approach:
- A series of standards are prepared.
- A calibration curve is generated.
- Unknown samples are compared against that curve.
While effective in many situations, external calibration assumes that standards and samples behave similarly during measurement. Groundwater samples can vary considerably in composition, potentially introducing matrix effects that influence analytical response. Standard addition minimizes these effects by adding known quantities of arsenic directly to the sample being analyzed.
As a result:
- Calibration occurs within the sample matrix.
- Matrix effects are partially compensated.
- Quantification is generally more robust.
12.3 The OpenArsenic standard addition scheme
The current OpenArsenic method uses:
Measurement 1: Sample
Measurement 2: Sample + 2.5 ppb arsenic
Measurement 3: Sample + 5 ppb arsenic
This workflow was selected during method development because it:
- Produces clear signal increases
- Remains within the validated working range
- Provides reliable regressions for typical groundwater samples
- Minimizes reagent consumption
12.4 Arsenic standard solution
The current OpenArsenic workflow uses: 1 ppm arsenic standard
This standard is used for all routine additions.
The standard should:
✔ Be prepared accurately
✔ Be clearly labeled
✔ Be protected from contamination
✔ Be replaced when quality is uncertain
The preparation of arsenic standards is described in Chapter 7.
12.5 What should happen during standard addition determination of arsenic
Following each standard addition, the arsenic peak should increase. Although replicate measurements will vary slightly, the overall trend should be clear.
In most successful measurements:
- Peak position remains similar.
- Peak shape remains similar.
- Peak height increases.

Figure 12.1. Example increase in arsenic signal following successive additions.
12.6 Arsenic calculation
After peak analysis is completed, OpenVoltam constructs a standard addition regression.
The regression uses:
- Added arsenic concentration
- Measured analytical signal
In the Sample Environment:
- Hold CTRL
- Select:
- Sample
- Standard Addition 1
- Standard Addition 2
- Click Calculate
OpenVoltam currently supports several calculation approaches. In the screen that opens, select peak height over baseline.
| Calculation approach | Explanation | Advantages | Disadvantages |
|---|---|---|---|
| Peak height over baseline (Recommended) | This method calculates peak height relative to the selected baseline. |
|
User-biases with baseline definition |
| Peak height from zero | Peak height is measured relative to the zero-current line. | Simple | More sensitive to baseline drift |
Derivative methods:
|
Peak height is not used to calculate arsenic concentration but the maximum slope of left-side, right-side or average is taken. |
|
|
The software then calculates the original arsenic concentration from the regression. For routine users, it is not necessary to perform these calculations manually. However, users should understand the basic principle.

Figure 12.2. Example regression generated from sample, +2.5 ppb, and +5 ppb measurements.
A successful regression should show:
✔ Increasing signal with increasing addition
✔ Approximately linear behavior
✔ No obvious outliers
If that is the case, select ‘Show results’ and the software will give the calculated arsenic concentration.

Figure 12.3: The calculated arsenic concentration
12.7 Warning signs
The following observations should prompt further investigation.
| Observation | Possible causes |
|---|---|
| No signal increase |
|
| Large scatter between replicates |
|
| Non-linear regression |
|
| Large confidence intervals |
|
12.8 Common mistakes
| Mistake | Recommendation |
|---|---|
| Incorrect standard concentration | Always verify the concentration of the arsenic standard. |
| Incorrect addition volume | Small pipetting errors can affect final calculations. |
| Poor documentation | Always record sample characteristics and additions clearly to avoid issues later on. |
| Changing multiple variables |
Do not change:
|
| Ignoring failed regressions | Poor regressions often indicate underlying analytical problems. Check your system. |