Chapter 13: Data acceptance criteria and quality control
This chapter provides practical guidance for evaluating data quality during routine arsenic monitoring.
13.1 Introduction
Producing an arsenic concentration is only part of the analytical process. Equally important is determining whether the result is trustworthy.
Every analytical method produces variability. Instrument noise, electrode condition, sample preparation, environmental conditions, and operator technique can all influence the final result. For this reason, the OpenArsenic workflow includes several quality-control checkpoints that help users determine whether a measurement should be accepted, repeated, or rejected.
This chapter provides practical guidance for evaluating data quality during routine arsenic monitoring. The criteria described here should be viewed as operational guidance developed during method development and user testing. As the OpenArsenic platform continues to evolve, acceptance criteria may be refined based on additional validation work and user experience.
13.2 Quality control philosophy
The OpenArsenic method relies on several independent indicators of data quality. A result should not be accepted solely because the software successfully calculated a concentration.
Instead, users should evaluate:
✔ Electrode performance
✔ Signal quality
✔ Replicate consistency
✔ Peak quality
✔ Standard addition response
✔ Regression quality
When multiple indicators suggest poor performance, the measurement should be investigated before reporting results.
13.3 Quality control checkpoints
The OpenArsenic workflow contains five major quality-control checkpoints.
- Checkpoint 1: Electrode conditioning
- Checkpoint 2: Raw voltammograms
- Checkpoint 3: Peak analysis
- Checkpoint 4: Standard addition regression
- Checkpoint 5: Final result review
13.4 Confidence intervals
OpenVoltam reports a confidence interval together with the arsenic concentration. Confidence intervals provide information about uncertainty in the calculated result.
In general:
- Smaller confidence intervals indicate greater confidence.
- Larger confidence intervals indicate greater uncertainty.
Large confidence intervals do not necessarily mean a result is incorrect. However, they often indicate weak signals, poor reproducibility or a large scatter between replicates.
Users should review the underlying data whenever unusually large confidence intervals are observed. Occasionally, a replicate may differ substantially from the others. Possible causes include electrode instability, air bubbles, electrical interference, user error or sample contamination.

Figure 13.1. Example of a replicate that may warrant further investigation.
Before excluding any data:
- Review the voltammogram.
- Review the peak analysis.
- Review notes from the measurement session.
- Determine whether a clear justification exists.
In the example from above, there is a clear outlier in the first standard addition run which may be caused by one of the reasons listed above. Because we have the luxury of running four replicates, we can simply exclude the outlier to achieve a reasonable R2 of 0.987.
In another example, we find that the signal does not change after the second addition (figure below). This most likely indicates that the operator forgot to add the standard after running the first standard addition. In this scenario, we can resolve the issue by performing the second standard addition and re-running of the measurement.

Figure 13.2. Example of a regression curve where the signal remains unchanged
13.5 Recommended routine quality-control practices
For routine groundwater monitoring:
Every Measurement Session
✔ Electrode conditioning
✔ Review of voltammograms
✔ Review of peak analysis
✔ Review of regression
Weekly
✔ Analyze a known arsenic standard
✔ Review reproducibility trends
Periodically
✔ Compare results with an independent analytical technique when available
✔ Review long-term instrument performance