Chapter 15: Method validation
Summary of validation work completed to date.
15.1 Introduction
Method validation is the process of demonstrating that an analytical method is fit for its intended purpose.
For the OpenArsenic platform, validation focuses on answering several key questions:
- Can the method reliably detect arsenic at environmentally relevant concentrations?
- Is the response linear across the concentration range of interest?
- Do measurements agree with established laboratory techniques?
- Can the method be used for routine groundwater monitoring?
- What are the current limitations of the method?
This chapter summarizes the validation work completed to date and places the OpenArsenic method within the broader context of arsenic voltammetry research.
It is important to note that OpenArsenic remains an actively developing platform. Validation studies continue as the method, software, and documentation evolve.
15.2 Validation objectives
The primary validation objectives of the OpenArsenic project were:
| Theme | Objectives |
|---|---|
| Analytical performance |
|
| Real water performance |
|
| Usability |
|
| Accessibility |
|
15.3 Scientific background
The OpenArsenic method did not emerge in isolation. The current platform builds upon more than a decade of research conducted by members of the project team and collaborators on electrochemical arsenic detection using gold microelectrodes and gold microwire electrodes.
Previous studies have demonstrated:
- Trace-level arsenic detection using gold microelectrodes
- Arsenic speciation using stripping voltammetry
- Nanomolar detection limits under optimized laboratory conditions
- Successful application of gold microwire electrodes to environmental samples
- Integration of low-cost open-source potentiostats into arsenic monitoring workflows
These studies provided the scientific foundation upon which the current OpenArsenic platform was developed.
Insert a historical schematic of all the voltammetry work on arsenic starting from 2007
Figure 15.1. Development of the OpenArsenic method builds upon previous research in gold electrode arsenic voltammetry.
15.4 Detection limit
Using the hardware and analytical method as described in this manual, the detection limit was assessed. One of the primary objectives of the project was achieving arsenic detection at concentrations relevant to drinking water monitoring.
Current validation data indicate a detection limit of approximately: 0.5 µg/L (0.5 ppb)
This detection limit is substantially below the World Health Organization drinking water guideline value of: 10 µg/L (10 ppb)
The achieved detection limit allows the method to:
- Detect concentrations below regulatory thresholds
- Evaluate treatment performance
- Monitor low-concentration groundwater systems
1. Perform several measurements (ideally 7–10) containing electrolyte + deionized water + low arsenic concentration (1 ppb)
2. Process each dataset to obtain peak heights
3. Calculate the standard deviation (σ) of the peak heights.
4. Determine the slope (m) of the calibration curve (standard additions or external standards)
5. Compute: LOD = 3 × σ / m
6. Report the LOD in µg L⁻¹ using the same units as your calibration.
15.5 Working range
The linear range tells us how far we can increase arsenic concentration before the relationship between signal and concentration begins to deviate from linearity as the working electrode gets saturated with analyte. Within this range, the sensor’s response follows a straight line (y = Sx + b), making quantification straightforward. Beyond it, the electrode surface becomes saturated, and the signal no longer grows proportionally. In the typical OpenArsenic method using the microwire electrode, we see a detection limit of ~0.3-0.5 ppb and linear range up to ~20-25 ppb, using a 20s deposition time.

Figure 15.2: Linear range of the OpenArsenic method
To determine the linear range, prepare a clean 0.1 M HCl electrolyte and ensure your blank is stable. Then start adding arsenic in small, known increments — for example, 2.5 ppb at a time — and record the corresponding stripping peak height for each addition.
Measure each addition three times to improve reliability and estimate experimental uncertainty. Once all data are collected, plot peak current (µA) against arsenic concentration (ppb) and include error bars representing the standard deviation of each triplicate.
Your calibration curve will initially show a straight line. At higher concentrations, you’ll notice that the slope begins to decrease — meaning that increases in arsenic concentration no longer produce proportionally larger peaks. This point marks the upper limit of your linear range.
For instance, in our example experiment, linearity was maintained up to approximately 24 ppb As. Beyond that, the curve started to flatten, indicating electrode surface saturation.
15.6 Groundwater validation study
The most important validation step involved comparison of OpenArsenic measurements with ICP-MS analysis using natural groundwater samples. A total of 11 groundwater samples were analyzed. These samples represented real groundwater matrices rather than laboratory-prepared standards. The objective was to determine whether OpenArsenic could accurately quantify arsenic in realistic environmental samples.
Each sample was analysed using the procedure described in this manual, with triplicate measurements at each standard addition step. Results were summarised as average concentration, standard deviation, and relative standard deviation (RSD). Where the RSD (Relative Standard Deviation) is calculated as:
RSD=StdevMeanx100% -— fix this formatting error
| Sample name | [As] (ppb) | Standard deviation (ppb) | Relative standard deviation (%) |
|---|---|---|---|
| Sample 1 | 2.14 | 0.10 | 4.8 |
| Sample 2 | 8.88 | 1.11 | 12.5 |
| Sample 3 | 12.56 | 1.35 | 10.8 |
| Sample 4 | 13.53 | 0.83 | 6.2 |
| Sample 5 | 17.04 | 0.66 | 3.9 |
| Sample 6 | 27.75 | 3.14 | 11.3 |
| Sample 7 | 8.58 | 0.24 | 2.8 |
| Sample 8 | 5.01 | 0.39 | 7.7 |
| Sample 9 | 17.20 | 1.78 | 10.4 |
| Sample 10 | 16.74 | 1.28 | 7.6 |
| Sample 11 | 12.71 | 1.12 | 8.8 |
Lower RSD values (typically <10%) indicate excellent precision for this system, with lower RSD achieved with more repeat scans or measuring sample multiple times.
Comparison to ICP-MS
Inductively Coupled Plasma Mass Spectrometry (ICP-MS) is widely regarded as one of the most sensitive and reliable techniques for trace metal analysis.
For each groundwater sample:
- OpenArsenic measurements were performed
- ICP-MS measurements were performed independently by an accredited external laboratory
- Results were compared
Overall, strong agreement was observed between the two methods (see figure 15.3). The comparison demonstrated that the OpenArsenic platform can provide arsenic concentrations comparable to those obtained using laboratory-based ICP-MS analysis.

Figure 15.3. Comparison of arsenic concentrations determined using OpenArsenic and ICP-MS.
15.7 Outliers and deviations
As with most environmental validation studies, not every sample exhibited identical agreement. One groundwater sample showed a larger deviation than the remaining dataset. Environmental samples can differ substantially in composition, and factors such as matrix effects, sample preservation, analytical variability and sample heterogeneity may contribute to differences between techniques.
Importantly, the overall agreement across the validation dataset remained strong despite the presence of individual deviations.
15.8 User validation
Analytical performance alone does not guarantee successful implementation of a new monitoring technology. For this reason, the OpenArsenic project includes a dedicated user validation component designed to evaluate whether new users can successfully operate the system, generate reliable results, and identify barriers to adoption.
The first round of user testing was conducted in April 2026 and involved five participants from two partner organizations: Caminos de Agua and Cantaro Azul. Participants represented a range of educational backgrounds and levels of laboratory experience, from routine water-quality practitioners to individuals with only limited prior analytical experience. The study was designed to assess usability, user confidence, trust in results, analytical performance, and the effectiveness of the existing documentation and software.
All participants attended a structured workshop consisting of three components:
- Introduction to arsenic contamination and the underlying measurement principles.
- Introduction to the required hardware, reagents, and instrumentation.
- A live demonstration of the complete measurement workflow.
Participants subsequently performed measurements with increasing levels of independence and completed questionnaires both before and after the workshop. The questionnaires assessed baseline knowledge, motivation, self-efficacy, system usability, confidence in results, procedural confidence, and willingness to use the system in real-world applications. Analytical performance was evaluated by comparing participant-generated arsenic concentrations against expert reference measurements.
Overall, the results were highly encouraging. Participants entered the workshop with high motivation and confidence in their ability to learn the method, regardless of their prior experience level. Following training, the OpenArsenic platform achieved a mean System Usability Scale (SUS) score of 70.5, exceeding the widely accepted benchmark score of 68 and placing the system in the "above average usability" category. Individual SUS scores ranged from 65.0 to 72.5.
Participants also reported high levels of trust in the analytical results, confidence in carrying out the measurement procedure, and willingness to use the system in future monitoring projects. Practical use intent scores were particularly strong, with all participants indicating that the system could be useful for community-based water quality monitoring and that the time required for a measurement was reasonable.
Analytical performance demonstrated that new users were capable of producing results that agreed reasonably well with expert reference measurements. Across the concentration range tested, most measurements fell within a ±30% agreement band relative to expert values. Performance was strongest at arsenic concentrations above 10 µg/L, where relative errors were generally below 10% for most participants. Importantly, the participant with the lowest educational background achieved the best overall analytical performance, demonstrating that prior scientific training is not necessarily required to operate the method successfully.
A consistent trend observed across all participants was that measurement errors increased at lower arsenic concentrations (<10 µg/L). This behaviour was independent of experience level and is believed to reflect the reduced signal-to-noise ratio inherent to electrochemical stripping voltammetry near the method detection limit rather than operator error. At concentrations above the World Health Organization drinking water guideline value of 10 µg/L, all participants achieved acceptable analytical performance, supporting the suitability of the OpenArsenic platform for groundwater monitoring and screening applications.
The user-testing study also provided valuable feedback for future development. The primary usability challenges identified by participants were related to software workflow, file management, and data processing rather than the analytical method itself. Common suggestions included improved software integration, automated peak detection, simplified file handling, automated data transfer between analysis steps, and clearer visual indicators of result quality. These findings directly informed the development of the OpenVoltam graphical user interface and the accompanying documentation.
Although this first user-testing round involved only five participants and primarily technically trained NGO staff, the results provide encouraging evidence that the OpenArsenic platform can be successfully transferred to new users with varying backgrounds. Additional beta testing, in which users operate the system independently using only the provided documentation and software, is planned to further evaluate the transferability and robustness of the method.
15.9 Current limitations
Although validation results are encouraging, several areas remain under active investigation.
Reproducibility: Long-term reproducibility studies are ongoing.
Current work includes:
- Multi-day measurements
- Operator comparisons
- Routine monitoring studies
Expanded matrix testing: Additional groundwater types may require future evaluation.
Long-term stability: Further work is required to characterize:
- Electrode lifetime
- Long-term performance
- Extended storage effects
Multi-site validation: Future testing at additional organizations will strengthen confidence in the method.
15.10 Strengths of the OpenArsenic platform
Current validation work demonstrates several important strengths.
| Strength | Explanation |
|---|---|
| Accessibility | The instrumentation is low-cost as compared to other arsenic detection tools offering similar reliability and reproducibility. |
| Open-source | The open-source nature of OpenArsenic makes it a completely transparent method. Users can inspect and understand the complete analytical workflow. |
| Sensitivity | Detection limit is well below the drinking water guideline values. LOD = 0.5 ppb while the WHO limit is 10 ppb (at the time of writing). |
| Agreement with laboratory methods | Strong agreement with ICP-MS as shown in Figure X. |
| Ease of deployment | Suitable for NGO and community-based monitoring applications. Hardware is portable and little laboratory equipment is needed. |