Chapter 4: Hardware assembly
This chapter describes how to assemble and connect the OpenArsenic hardware.
4.1 Introduction
This chapter describes how to assemble and connect the OpenArsenic hardware. The assembly process typically takes less than 15 minutes once users are familiar with the system.
✔ Install OpenVoltam
✔ Verify all hardware is available
✔ Inspect cables and electrodes
✔ Prepare a clean workspace
4.2 Hardware layout
Arrange the following components: rodeostat, computer, electrochemical cell, electrodes, magnetic stirrer and the relay. The setup should allow easy access to all components while minimizing cable strain.

Figure 4.1. Example arrangement of OpenArsenic hardware components.
4.3 Connecting the rodeostat and the electrodes
Connect the Rodeostat to the computer using the USB cable provided. Connect the electrodes to the Rodeostat according to the designated terminals using the screwdriver.
- Working electrode: Connect the gold microwire to the working electrode terminal (WRK)
- Reference electrode: Connect the Ag/AgCl reference electrode to the reference terminal (REF)
- Counter electrode: Connect the counter electrode to the counter terminal (CTR)
Ensure all connections are secure before beginning measurements.

Figure 4.2: Rodeostat and electrodes connections
4.5 Installing the electrodes in the cell
Place the electrodes into the electrochemical cell. It is recommended to place the working electrode near the centre, the reference electrode close to the working electrode and the counter electrode away from the working electrode.
✔ Electrodes touching each other
✔ Electrodes touching the stir bar
✔ Excessive movement during measurements

Figure 4.3: Recommended electrode positioning
4.6 Installing the relay
If you are not using a relay, the stirrer can still be controlled manually. However, automatic relay control is strongly recommended because it improves consistency and reduces operator workload.

Figure 4.4: Installing the relay expansion board, cables and connecting to the rodeostat and stirrer
Step 1 – Install the relay expansion board
Attach the supplied relay expansion board to the top of the Rodeostat using the mounting screws provided. Ensure that the board is firmly attached and does not move during operation.
Step 2 – Connect the Ribbon Cable
Locate the DIO (Digital Input/Output) header on the Rodeostat. Connect the wide ribbon cable between:
- The DIO header on the Rodeostat
- The matching connector on the relay expansion board
Verify that the cable is fully inserted and correctly oriented before proceeding.
Step 3 – Connect the Relay Control Wires
Connect the two control wires from the relay to the relay expansion board. Positive (+) goes to the DIO3 pin, negative (-) goes to GND (next to DIO3).
Step 4 – Connect the magnetic stirrer
Insert the power plug of the magnetic stirrer into the relay outlet labeled: "Normally Off"
This ensures that the stirrer only receives power when activated by OpenVoltam. Do not connect the stirrer to an outlet labeled "Always On", as this would prevent automatic control.
Step 5 – Power the relay
Connect the relay to a mains power outlet. Switch the relay on using the power switch located on the relay housing. Verify that the relay power indicator illuminates.

Figure 4.5: The complete hardware setup of OpenArsenic
4.7 Installing the stir bar
Place the stir bar into the electrochemical cell before adding solution. Verify it can rotate freely, that stirring is stable and that the stir bar does not collide with the electrodes.