Chapter 11: Arsenic measurement procedure
This chapter describes the standard procedure used to measure dissolved arsenic using the OpenArsenic platform.
11.1 Introduction
This chapter describes the standard procedure used to measure dissolved arsenic using the OpenArsenic platform.
The procedure outlined here represents the validated OpenArsenic workflow and should be followed whenever possible. Deviations from the procedure should be documented and may require additional validation.
The OpenArsenic method combines:
- Electrochemical arsenic detection
- Standard addition calibration
- Replicate measurements
- Peak analysis
- Statistical evaluation
Together, these steps provide a robust framework for quantifying arsenic in groundwater samples.
This chapter should be considered the primary Standard Operating Procedure (SOP) for routine arsenic analysis using OpenArsenic.
11.2 Overview of the workflow
A typical arsenic measurement consists of:

Figure 2. Sequence of steps used for routine arsenic analysis.
11.3 Before you begin
Before starting a measurement, verify:
| Hardware | Reagents | Electrode | Software |
|---|---|---|---|
|
✔ Rodeostat connected
✔ Electrodes installed
✔ Stirrer operational
✔ OpenVoltam running
|
✔ Sample prepared
✔ 1 M HCl available
✔ Arsenic standard available
✔ Deionized water available
|
✔ Conditioned today
✔ No visible contamination
✔ Stable conditioning response
|
✔ Sample environment created
✔ Arsenic Detection method available
|
11.4 Creating the sample run
- Open the desired sample environment.
- Select: New Run
- And follow the recommendations in the figure below
• Better reproducibility estimates
• Improved confidence intervals
• Identification of outliers
• Increased confidence in final results

Figure 11.1. Configuration of the initial sample measurement.
11.5 Running the measurement
Select: Ready to Run
Review the settings carefully. Then select: Begin Run
✔ Observe the live voltammogram
✔ Verify normal software behavior
✔ Verify stable hardware operation
✔ Confirm stirrer operation during deposition
Avoid:
✘ Disconnecting the USB cable
✘ Moving electrodes
✘ Closing OpenVoltam
✘ Running resource-intensive software

Figure 11.2. Example live voltammogram displayed during measurement.
At this stage, focus on identifying obvious problems such as:
• Saturation
• Excessive noise
• Missing signal
• Communications errors
11.6 Peak analysis
When the run finishes:
- Data are automatically stored
- Voltammograms are saved
- Results become available for analysis
At this stage users may:
- Analyze the data immediately
- Continue with additional measurements
- Return later for analysis

When analysis begins, OpenVoltam automatically estimates:
- Peak position
- Baseline location
- Start point
- End point
This automation helps:
- Reduce operator variability
- Increase consistency
- Speed up analysis
For the current Caminos de Agua arsenic method, peaks typically occur between: -0.2 V and +0.1 V
However, this range may vary depending on electrode condition and sample matrix. After opening the analysis window:
- Review the automatically selected baseline.
- Confirm that the full arsenic peak is included.
- Verify that the baseline does not intersect the peak.
- Check that the baseline extends beyond the peak region on both sides.

Adjusting the Baseline
If the automatically selected baseline is not appropriate:
- Drag the first baseline marker to a more suitable location.
- Press Z to move to the second baseline marker.
- Adjust the second marker if necessary.
- Verify that the baseline follows the underlying signal rather than the arsenic peak itself.
Once satisfied: click Accept & Next and continue to the next repetition. After all repetitions have been analyzed:
- Click Save All
- Return to the Sample Environment
11.7 Performing standard addition 1 and 2
- Once the sample measurement is complete, right-click the Sample run and select: New Run From Config
- Change the Run Type from Sample to Standard Addition
- Enter: Volume of arsenic standard added and Concentration of the added standard
Notes 🡺 Example: Well_A + 2.5 ppb As

Selecting an appropriate spike
The first standard addition should produce a noticeable increase in signal while remaining within the linear working range of the method.
General recommendations:
- Avoid extremely small additions.
- Avoid excessively large additions.
- Aim for a clear increase in peak height: we recommend adding 2.5 ppb arsenic.
Examples and further guidance are provided in Chapter 12.
Running the standard addition measurements
- Repeat the measurement using the same number of repetitions as the original sample.
- Recommended: Four repetitions
- Verify that acquisition proceeds normally.
- When complete, review the voltammograms for obvious issues and analyse the peaks.
🡺 do the same for the second standard addition
11.8 Common problems during data acquisition
| Issue | Possible causes |
|---|---|
| No peak visible |
|
| Excessive noise |
|
| Unstable signal |
|
| Saturation |
|