Part 17. Best practices for routine use
List of best practices and recommended daily routine.
- Use the preconfigured methods for routine arsenic measurements.
- Maintain consistent electrode preparation.
- Use four repetitions per run.
- Analyze peaks immediately after acquisition when possible.
- Use consistent approach to baseline placement.
- Keep clear run names and notes.
- Avoid multitasking during measurements.
- Inspect regression linearity before accepting results.
- Use New Run from Config to save time and reduce setup errors.
- Record any unusual observations in the notes field or lab notebook.
Recommended daily routine
| Stage | Recommended action |
|---|---|
| Start of day | Inspect electrodes, prepare workspace, open OpenVoltam, create/open sample workspace. |
| Before samples | Condition electrode, run blank or standard check if required. |
| During analysis | Run sample, standard addition 1, standard addition 2, with four repetitions each. |
| After each run | Analyze peaks immediately, setup and save calculation. |
| End of day | Export or back up data, rinse electrodes, store solutions and electrodes properly, dispose/collect waste. |